In thermal ionization, also referred to as surface ionization, chemically-purified material loaded onto a filament which is then heated to cause some of the material to be ionized as it boils off the hot filament. Filaments are generally flat pieces of metal around 1-2mm wide, 0.1mm thick, bent into an upside-down U shape and welded to steel posts that supply a current.

Saha-Langmuir equation

The likelihood of ionisation is a function of the filament temperature, the work function of the filament substrate and the ionization energy of the element.

This is summarised in the Saha-Langmuir equation:[1]

<math>\frac{Y_1}{Y_0} = \frac{g_1}{g_0} \exp \Bigg(\frac{\phi-IP}{kT}\Bigg)</math>
<math>\frac{Y_1}{Y_0}</math> = ion to neutral ratio
<math>\frac{g_1}{g_0}</math> = statistical weights of ion and neutral states
<math>\phi</math> = surface work function
IP = element ionization potential
k = Boltzmann's constant
T = surface temperature

Thermal ionization mass spectrometry

One application of thermal ionization is thermal ionization mass spectrometry (TIMS). The ions being produced at the filament are directed into a mass spectrometer to analyze the elements or isotopes present in the sample.[2]